Zeiss LSM 510 META NLO
LSM unit
- Scanning method: two galvanometer scanning mirrors
- Standard detection via 2 photomultiplier tubes (PMT Ch1 and 2) each equipped with 8 position filter wheels.
- Fluorescence can be spectrally dispersed and measured via spectral detector ChS (META detector, 32 PMT array)
- Each PMT covers 10.7 nm wavelength of spectrum
- Spectral bandwidth: 340 nm (380 - 720 nm)
- 8 virtual channels ChS1-8 can be monitored simultaneously
- A transmission detector (PMT ChD) enables bright field/DIC imaging using the scanning laser
- A non-descanned detector can be used for two-photon imaging
- Scanning modes: X, Y, Z, t and λ, uni- or bidirectional scanning
- Image resolution: max. 2048 x 2048
- Scanning speed: up to 5 frames/s at 512 x 512 pixels
- Dynamic range: 12-bit for each PMT
- Zoom: 1x, ..., 40x; continuous control
- Fully motorized pinholes for each PMT (Ch1, 2, S): displaceable in x,y,z-direction and diameter