Main Applications:
3d/4d cLSM
Person in Charge:
Viktoria Liss
Location:
Building 37, Room 106,
S2-laboratory
Leica LSM SP5
LSM unit
- Scanning method: two galvanometer scanning mirrors
- Excitation and emission light is separated via different polychroic beamsplitters (6 position filter wheel, see Filter)
- Fluorescence detection via 3 independent low-noise photomultiplier tubes
- Fluorescence signal is spectrally dispersed and guided via the spectrophotometric detector to the photomultiplier tubes
- A transmission detector enables bright field/DIC imaging using the scanning laser
- Scanning modes: X, Y, Z, time t and wavelength λ, uni- or bidirectional scanning
- Image resolution: max. 8192 x 8192
- Scanning speed: up to 5 frames/s at 512 x 512 pixels
- Hardware zoom: 1x, ..., 64x; continuous control